Mask induction despite circuit obstruction: an unrecognized hazard of relying on automated machine check technology.

Autor: Yang KK; From the Department of Pediatric Anesthesia, C.S. Mott Children's Hospital, University of Michigan, Ann Arbor, Michigan., Lewis IH
Jazyk: angličtina
Zdroj: A & A case reports [A A Case Rep] 2014 Jun 15; Vol. 2 (12), pp. 143-6.
DOI: 10.1213/XAA.0000000000000026
Abstrakt: Various equipment malfunctions of anesthesia gas delivery systems have been previously reported. Our profession increasingly uses technology as a means to prevent these errors. We report a case of a near-total anesthesia circuit obstruction that went undetected before the induction of anesthesia despite the use of automated machine check technology. This case highlights that automated machine check modules can fail to detect severe equipment failure and demonstrates how, even in this era of expanding technology, manual checks still remain essential components of safe care.