Comments on ;A cone-beam filtered backprojection reconstruction algorithm for cardiac single photon emission computed tomography' by G. T. Gullberg and G. L. Zeng.
Autor: | Wang G; Dept. of Electr. & Comput. Eng., State Univ. of New York, Buffalo, NY., Lin TH |
---|---|
Jazyk: | angličtina |
Zdroj: | IEEE transactions on medical imaging [IEEE Trans Med Imaging] 1993; Vol. 12 (2), pp. 383-4. |
DOI: | 10.1109/42.232270 |
Abstrakt: | In the above-titled work by G. T. Gullberg and G. L. Zeng (ibid., vol.11, no.1, p.91-101, 1992), a fan-beam reconstruction formula of a noncircular scanning locus was derived and extended for half-scan cone-beam reconstruction. However, the reconstruction formula is not exact mathematically unless a necessary condition is satisfied. In this correspondence, the commenters derive this necessary condition and provide a geometrical explanation of the condition. |
Databáze: | MEDLINE |
Externí odkaz: |