Improvement of visualization in computed tomographic colonography after mechanic air insufflation.
Autor: | Sirakov NV; Department of Imaging Diagnostics, Nuclear Medicine and Radiation Therapy, Medical University, Plovdiv, Bulgaria. mail@nsirakov.com, Velkova KG, Nikolov RR, Sirakov VN |
---|---|
Jazyk: | angličtina |
Zdroj: | Folia medica [Folia Med (Plovdiv)] 2006; Vol. 48 (1), pp. 46-9. |
Abstrakt: | Objective: Our aim was to develop an insufflation system for CT colonography enabling even and quick air inflation as well as pressure control and optional emergency decompression. Patients and Methods: A modified air insufflation system for CT colonography, consisting of an electric air pump, manometer, safety valve, and a balloon tip is presented. Thirty one patients underwent CT colonography from August to December, 2004. Results: Optimal distension was achieved in 30 patients; the distension was incomplete (partial) only in one case. The insufflation was performed within 60 seconds in 28 of the cases, in 2 patients--within 180 seconds and in one patient--after 180 seconds. Conclusions: The presented insufflation system allows optimization of CT colonography technique, reduces the risk of adverse events during examination and improves the precision of the method. |
Databáze: | MEDLINE |
Externí odkaz: |