Autor: |
Ross AS; Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY 12180-3590, USA. rossa@rpi.edu, Saulnier GJ, Newell JC, Isaacson D |
Jazyk: |
angličtina |
Zdroj: |
Physiological measurement [Physiol Meas] 2003 May; Vol. 24 (2), pp. 509-16. |
DOI: |
10.1088/0967-3334/24/2/361 |
Abstrakt: |
Questions regarding the feasibility of using electrical impedance tomography (EIT) to detect breast cancer may be answered by building a sufficiently precise multiple frequency EIT instrument. Current sources are desirable for this application, yet no current source designs have been reported that have the required precision at the multiple frequencies needed. We have designed an EIT current source using an enhanced Howland topology in parallel with a generalized impedance converter (GIC). This combination allows for nearly independent adjustment of output resistance and output capacitance, resulting in simulated output impedances in excess of 2 Gohms between 100 Hz and 1 MHz. In this paper, the theoretical operation of this current source is explained, and experimental results demonstrate the feasibility of creating a high precision, multiple frequency, capacitance compensated current source for EIT applications. |