Zobrazeno 1 - 10
of 48
pro vyhledávání: '"William B. Handler"'
Publikováno v:
Magnetic Resonance in Medicine. 87:3011-3021
Delta relaxation-enhanced MR (dreMR) is a field-cycling quantitative method for molecular imaging. The dreMR method uses a BA solenoid produces the bulk field shift, and a boundary element method is employed to design in-series shim and shield layers
Publikováno v:
Brain and Mind Institute Researchers' Publications
In this article, an ultralow-frequency electric field probe capable of measuring time-varying gradient coil induced electric fields in a tissue-mimicking material was constructed, calibrated, and validated. This probe consisted of a 2 cm short dipole
Publikováno v:
Brain and Mind Institute Researchers' Publications
This paper presents the design and validation of a tuned time-domain electric field probe for mapping of radio frequency (RF) exposures used during testing of magnetic resonance imaging (MRI) conditional medical devices. The probes were 5 and 10 mm s
Publikováno v:
Brain and Mind Institute Researchers' Publications
In this work a 21 MHz radiofrequency (RF) exposure system was designed, constructed and evaluated as a testing platform for MRI device safety. The exposure system consisted of a birdcage coil powered in quadrature by a power system consisting of a hy
Publikováno v:
Brain and Mind Institute Researchers' Publications
CCECE
CCECE
As a part of this work a small patch antenna probe was developed to measure the variation in the electric field produced by gradient coils within an MRI in the presence of any active implantable medical devices (AIMDs). This probe was designed, fabri
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::639a0750fa2e174d4bfd3e2625b905da
https://ir.lib.uwo.ca/brainpub/727
https://ir.lib.uwo.ca/brainpub/727
Autor:
Ryan Chaddock, L. Martyn Klassen, Blaine A. Chronik, Brian Dalrymple, Justin Peterson, William B. Handler, Joseph S. Gati, Geron Andre Bindseil, Chad T. Harris, Kyle M. Gilbert, Frank Van Sas
Publikováno v:
Brain and Mind Institute Researchers' Publications
A gradient coil with integrated second and third order shims has been designed and constructed for use inside an actively shielded 310 mm horizontal bore 9.4 T small animal MRI. An extension of the boundary element method, to minimise the power depos
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::90317a4569676ba7c4de92b72c9484dd
https://ir.lib.uwo.ca/brainpub/719
https://ir.lib.uwo.ca/brainpub/719
Publikováno v:
Brain and Mind Institute Researchers' Publications
© 2018 IEEE. This paper presents the design and construction of a magnetic field probe with an active balun for conditional testing of medical devices within a magnetic resonance imaging (MRI). The magnetic field probe was designed using a small loo
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::43d19a65606c0293fdb703eb852901e6
https://ir.lib.uwo.ca/brainpub/383
https://ir.lib.uwo.ca/brainpub/383
Reliable RF B/E-Field Probes for Time-Domain Monitoring of EM Exposure During Medical Device Testing
Publikováno v:
Brain and Mind Institute Researchers' Publications
This paper presents electric and magnetic probes to measure radio frequency (RF) electric and magnetic fields for measuring the time-varying RF fields used in magnetic resonance imaging (MRI). A small single loop (2 cm) and short dipole antenna (2 cm
Publikováno v:
Electronics Letters. 54:16-18
A new RF B/H field probe for time-domain monitoring of RF electromagnetic-radiated exposure during medical device testing inside the lossy tissue-mimicking material is presented. The probe was tuned and matched for a centre frequency of 127.6 MHz to
Autor:
Ryan Chaddock, Arjama Halder, William B. Handler, Blaine A. Chronik, Ali Attaran, Paul Christiaans
Publikováno v:
CCECE
In this work ultra-low frequency probes for measuring time-varying electric field produced in gradient coils were designed, fabricated and tested. A 2 cm short dipole antenna followed by an instrumentation amplifier was chosen to measure the electric