Zobrazeno 1 - 10
of 40
pro vyhledávání: '"Torsten Funck"'
Publikováno v:
IEEE Transactions on Instrumentation and Measurement. 68:2003-2006
The sensitivity of planar multijunction thermal converters was significantly increased by filling the housing with noble gasses, preferably xenon or krypton. The performance was observed for almost two years, and no degradation of the sensitivity was
Publikováno v:
Measurement Science and Technology. 33:085016
Impedance standards having impedance values in the low mΩ range, with non-zero reactances, and operable with alternating current-currents of a few Amps and up to a few kHz are needed for appropriate impedance meter calibration. These challenging ope
Publikováno v:
2020 Conference on Precision Electromagnetic Measurements (CPEM).
Inductance standards currently available are made up of toroidal inductors, i.e., coils of wire wound around a toroidal nonmagnetic core. The maximum inductance achievable is 10 H, while modern LCR meters offer measurement ranges up to more than 10 k
Publikováno v:
2020 Conference on Precision Electromagnetic Measurements (CPEM).
AC–DC transfer is performed almost exclusively using thermal converters. At voltages above a few volts, series resistors are used to extend the permissible voltage, which leads to significant power being drained from the source. By contrast, sensin
Autor:
Thomas Spiegel, Torsten Funck
Publikováno v:
IEEE Transactions on Instrumentation and Measurement. 66:1411-1416
Traceability for ac currents below 1mA is needed to calibrate calibrators and multimeters at frequencies up to at least 100 kHz. Since the sensitivity of thermal convertors is not sufficient, only transfer standards using amplifiers are suitable. A n
Autor:
Thomas Spiegel, Torsten Funck
Publikováno v:
IEEE Transactions on Instrumentation and Measurement. 66:1454-1458
Disk resistors are widely used in micropotentiometers up to frequencies in the radio frequency range. They offer very small AC–DC differences but the permissible power dissipation is quite small. Also, the resistance value is poorly defined due to
Publikováno v:
IEEE Transactions on Instrumentation and Measurement. 64:1600-1605
A determination of the frequency behavior of the Physikalisch-Technische Bundesanstalt (PTB) calculable capacitor in the audio frequency range is carried out by applying an equivalent circuit model. The circuit model considers the influence of distri
Publikováno v:
Funck, T, Bothe, H, Kölling, A, Satrapinski, A & Suomalainen, E-P 2017, ' Bilateral international inductance comparison using traveling standards in thermoregulated enclosures ', IEEE Transactions on Instrumentation and Measurement, vol. 66, no. 6, 7801843, pp. 1511-1515 . https://doi.org/10.1109/TIM.2016.2634720
Funck, T, Bothe, H, Kolling, A, Satrapinski, A & Suomalainen, E-P 2016, Bilateral international inductance comparison using traveling standards in thermoregulated enclosures . in 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016) ., 7540510, IEEE Institute of Electrical and Electronic Engineers, Conference on Precision Electromagnetic Measurements, CPEM 2016, Ottawa, Canada, 10/07/16 . https://doi.org/10.1109/CPEM.2016.7540510
Funck, T, Bothe, H, Kolling, A, Satrapinski, A & Suomalainen, E-P 2016, Bilateral international inductance comparison using traveling standards in thermoregulated enclosures . in 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016) ., 7540510, IEEE Institute of Electrical and Electronic Engineers, Conference on Precision Electromagnetic Measurements, CPEM 2016, Ottawa, Canada, 10/07/16 . https://doi.org/10.1109/CPEM.2016.7540510
Standards with toroidal windings are traditionally used for international comparisons of inductance. The achieved uncertainties of the comparisons are dominated by the rather high temperature coefficient of the traveling standards. Hence, thermoregul
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::c595069ad987ed3eed60ac10c0f46fdc
https://doi.org/10.1109/CPEM.2016.7540510
https://doi.org/10.1109/CPEM.2016.7540510
Autor:
Thomas Spiegel, Torsten Funck
Publikováno v:
2016 Conference on Precision Electromagnetic Measurements (CPEM 2016).
Calibrators are compact devices generating reference values for voltage, current, resistance and other quantities and are mainly used to calibrate measuring devices. Especially for AC quantities, there are wide parameter ranges for the generated valu
Autor:
Torsten Funck, Thomas Spiegel
Publikováno v:
2016 Conference on Precision Electromagnetic Measurements (CPEM 2016).
Traceability for AC currents below 1 mA is needed to calibrate calibrators and multimeters at frequencies up to at least 100 kHz. Since the sensitivity of thermal convertors is not sufficient, only transfer standards using amplifiers are suitable. A