Zobrazeno 1 - 6
of 6
pro vyhledávání: '"T. W. Fengler"'
Publikováno v:
Surgical Endoscopy. 14:388-394
On the basis of experience gained from 6,000 laparoscopies (73% cholecystectomies) at the Moabit Hospital in Berlin, we carried out a cohort study to analyze the failure rate and decontamination of labeled "tracer" instruments processed in three test
Publikováno v:
Surgical Endoscopy. 12:1275-1279
Because so many common surgical problems can now be addressed by the laparoscopic approach, the issue of sterile processing has to be reconsidered.Selected laparoscopic instrumentation was analyzed regarding wear and tear and decontamination after st
Publikováno v:
Minimally Invasive Therapy. 4:203-209
SummarySelected optical, mechanical, puncturing, electromechanical instruments and accessories for laparoscopic surgery were studied to obtain data regarding repair and cleaning modalities. The instruments were used for more than 3000 laparoscopies a
Publikováno v:
Materials for Medical Engineering, Volume 2
* D 33332 Gutersloh, GERMANY Abstract: Tracer instruments (in vivo), test soils and test probes (in vitro) are needed to examine surface layers which might contain infective material. Disinfection efficacy with the reduction of colony forming units (
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_________::be252cd63b3bcf9a7552e17107c2090c
https://doi.org/10.1002/3527606149.ch26
https://doi.org/10.1002/3527606149.ch26
Publikováno v:
Klinik und Forschung in der Chirurgie unter dem Aspekt von Effizienz und Ökonomie ISBN: 9783540633327
Wear, tear, and possible contamination after 100 clinical cycles of three test trays were investigated on laparoscopic “tracer” instruments in a prospective clinical study. Failure was below 4 per 100 laparoscopies, and remaining contamination wa
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_________::49630969768f1d78deba20d95806d441
https://doi.org/10.1007/978-3-642-60881-0_312
https://doi.org/10.1007/978-3-642-60881-0_312
Publikováno v:
Langenbecks Archiv fur Chirurgie. Supplement. Kongressband. Deutsche Gesellschaft fur Chirurgie. Kongress. 114
Wear, tear, and possible contamination after 100 clinical cycles of three test trays were investigated on laparoscopic "tracer" instruments in a prospective clinical study. Failure was below 4 per 100 laparoscopies, and remaining contamination was de