Zobrazeno 1 - 10
of 13
pro vyhledávání: '"Sofie De Cooman"'
Autor:
Louise Cuveele, Jan F. A. Hendrickx, Andre M. De Wolf, Sofie De Cooman, Brian B. Chesebro, Jeffrey Feldman, Jodi Sherman
Publikováno v:
Journal of Clinical Monitoring and Computing. 37:227-235
Autor:
Louise, Cuveele, Jan F A, Hendrickx, Andre M, De Wolf, Sofie, De Cooman, Brian B, Chesebro, Jeffrey, Feldman, Jodi, Sherman
Publikováno v:
Journal of clinical monitoring and computing.
Anesthetic agent consumption is often calculated as the product of fresh gas flow (FGF) and vaporizer dial setting (F
Autor:
Mohab Omer, Jan F. A. Hendrickx, Simon De Ridder, Alexander De Houwer, Rik Carette, Sofie De Cooman, Andre M. De Wolf
Publikováno v:
Journal of Clinical Monitoring and Computing. 32:799-806
Autor:
Sofie De Cooman, Jan F. A. Hendrickx, Simon P. A. J. De Ridder, Rik Carette, Andre M. De Wolf, Alexander Dehouwer
Publikováno v:
Journal of Clinical Monitoring and Computing. 30:193-202
Low flow anesthesia increases the use of CO2 absorbents, but independent data that compare canister life of the newest CO2 absorbents are scarce. Seven different pre-packed CO2 canisters were tested in vitro: Amsorb Plus, Spherasorb, LoFloSorb, Medis
Autor:
Idris E. Ghijselings, Philip J Peyton, Sofie De Cooman, Jan F. A. Hendrickx, Andre M. De Wolf, Rik Carette
Publikováno v:
Journal of Clinical Monitoring and Computing. 30:63-68
Current hypoxic guards systems fail to maintain the inspired O2 concentration (FIO2) ≥ 21 % across the entire fresh gas flow (FGF) range when a second carrier gas is used (N2O or air). We examined the performance of the Maquet O2 Guard(®), a smart
Autor:
Caroline Schollaert, Tom C. R. V. Van Zundert, Andre M. De Wolf, Philip J Peyton, Sofie De Cooman, Jan F. A. Hendrickx
Publikováno v:
Journal of Clinical Monitoring and Computing. 29:491-497
Because a case report and theoretical mass balances suggested that hypoxic guard systems may not prevent the formation of hypoxic inspired mixtures (FIO2 ≤ 21 %) over the clinically used fresh gas flow (FGF) range, we measured FIO2 over a wide rang
Autor:
Jan F A, Hendrickx, Simon P A J, De Ridder, Alexander, Dehouwer, Rik, Carette, Sofie, De Cooman, Andre M, De Wolf
Publikováno v:
Journal of clinical monitoring and computing. 30(2)
Low flow anesthesia increases the use of CO2 absorbents, but independent data that compare canister life of the newest CO2 absorbents are scarce. Seven different pre-packed CO2 canisters were tested in vitro: Amsorb Plus, Spherasorb, LoFloSorb, Medis
Publikováno v:
BMC Research Notes
Background Earlier software versions of the Zeus® (Lübeck, Dräger, Germany) failed to provide true closed circuit anesthesia (CCA) conditions. We examined whether the latest software (SW 4.03 MK 04672–00) achieves this goal. Methods In 8 ASA I
Autor:
José Coddens, Andre M. De Wolf, Thierry Deloof, D. Vandeput, Sofie De Cooman, Jan F. A. Hendrickx
Publikováno v:
Anesthesia & Analgesia. 93:391-395
UNLABELLED We determined the performance of the vaporizer of the ADU machine (Anesthesia Delivery Unit; Datex-Ohmeda, Helsinki, Finland). The effects of carrier gas composition (oxygen, oxygen/N(2)O mixture, and air) and fresh gas flow (0.2 to 10 L/m
Publikováno v:
BMC Anesthesiology
BMC Anesthesiology, Vol 12, Iss 1, p 22 (2012)
BMC Anesthesiology, Vol 12, Iss 1, p 22 (2012)
Background Hyperventilation may be used to hasten recovery from general anesthesia with potent inhaled anesthetics. However, its effect may be less pronounced with the newer, less soluble agents, and it may result in rehypnotization if subsequent hyp