Publikováno v:
American Journal of Clinical Pathology. 84:518-523
The authors evaluated the Kodak Ektachem Slides for total, "direct," conjugated, unconjugated, and albumin-bound ("delta") bilirubin. For various concentrations of control material, the precision (CV) within- and between-day ranged from 1.7 to 2.4% (
Publikováno v:
Diabetes Care. 9:189-193
We have evaluated a new commercially available radioimmunoassay kit for albumin determination in urine. The assay is precise; within-run precision (CV) in the clinically significant ranges is 1.8-3.5%, between-run, 1.2-8.5%. The minimum detection lim
Publikováno v:
Clinical Biochemistry. 7:206-211
Five commercial radioimmunoassay kits for the analysis of serum or plasma digoxin levels were evaluated. The adequacy of the literature supplied with the kits, within-the-day precision, recovery of added digoxin, clinical interpretation of results, a
Publikováno v:
Clinical Biochemistry. 17:15-18
We evaluated the Kodak Ektachem multilayer ammonia and creatinine slides. For various concentrations of control material the precision (CV) within- and (between-day) ranged from 0.8 to 7.8% (4.4-9.0%) for ammonia, and 1.9-25.1% (3.6-13.9%) for creati
Publikováno v:
Clinical Chemistry. 26:133-137
Some basic laboratory performance criteria were studied for the Eastman Kodak glucose and urea nitrogen analyzer. Serum, plasma, urine, and cerebrospinal fluid were tested. Precision, both "within-day" and "between-day," for both analytes was less th
Publikováno v:
Clinical Chemistry. 29:129-132
We evaluated the Kodak Ektachem multilayer ion-selective electrolyte slides. For various types and concentrations of control material the precision (CV) within- and between-day ranged from 0.5 to 1.3% (1.7-2.1%) for sodium, 1.2 to 2.2% (2.7-2.9%) for
Publikováno v:
Clinical Chemistry. 21:1922-1926
We evaluated six commercial radioassay kits in which tritium or iodine tracers are used to measure serum folate. We assessed the adequacy of the literature supplied with the kits, within-run precision, reagent stability, analytical recovery of added
Autor:
Norman P. Kubasik, Jagdish C. Mangla, Michael D. Turner, Gilda Guarasci, Eric A. Schenk, L. Desbaillets
Publikováno v:
Gastroenterology. 70:669-676
Four cases of Barrett's esophagus are presented. Three cases presented with significant esophageal bleeding and one case presented with high esophageal stricture. Gastrointestinal panendoscopy was done in each case and multiple biopsies were taken. T
Publikováno v:
Clinical Biochemistry. 5:266-270
Autor:
Michael T. Volosin, Norman P. Kubasik
Publikováno v:
Clinical Chemistry. 19:954-958
Lead, cadmium, and thallium are determined in a single 3-ml urine sample. The three metals, simultaneously chelated with sodium diethyldithiocarbamate, are extracted (at pH 3-7) into methyl isobutyl ketone (isopropylacetone), and the carbon rod atomi