Zobrazeno 1 - 6
of 6
pro vyhledávání: '"Marieke Verstraete"'
Publikováno v:
Journal of Analytical Atomic Spectrometry. 17:897-903
In this work, a method has been developed for the fast and reliable determination of silicon in polyamides (two samples with silicon contents of 10 and 50 µg g−1 were analyzed) by means of solid sampling electrothermal vaporization ICP-MS. For all
Publikováno v:
Journal of Analytical Atomic Spectrometry. 16:793-800
It is known that the accurate determination of ppm levels of sulfur in solid samples is very complicated. One of the approaches that have been evaluated in order to improve the detection limits for this element is the use of electrothermal vaporizati
Publikováno v:
Journal of Analytical Atomic Spectrometry. 16:1018-1027
Electrothermal vaporization (ETV) ICP mass spectrometry is a method that combines the ability of the graphite furnace to handle complex samples with the detection power of ICP-MS. It is somewhat surprising, however, that most works reporting on the a
Publikováno v:
Analytical Chemistry. 72:4310-4316
Next to laser ablation (LA) also electrothermal vaporization (ETV) from a graphite furnace as a means of sample introduction opens possibilities for direct analysis of solid samples using inductively coupled plasma mass spectrometry (ICPMS). In this
Autor:
Luc Moens, André van Alphen, Eric Denoyer, Marieke Verstraete, Martín Resano, Frank Vanhaecke
Publikováno v:
Journal of Analytical Atomic Spectrometry. 15:389-395
The coupling of a graphite furnace to an ICP mass spectrometer results in a method that combines the suitability of the graphite furnace for handling solid samples with the detection power of ICP mass spectrometry (ICP-MS). Surprisingly, the majority
Publikováno v:
Analytical Communications. 36:89-92
Electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICPMS) was used for the determination of the Pd content in a solid aliphatic polyketone. The presence of Pd in this material has to be attributed to the use of a Pd-contain