Zobrazeno 1 - 10
of 26
pro vyhledávání: '"Kyung Am Park"'
Publikováno v:
Flow Measurement and Instrumentation. 22:402-405
A Critical Flow Venturi Nozzle (CFVN) is usually calibrated using dry air. Yet CFVNs in industrial and calibration service centers are often used to measure flow rates of humid air. Therefore, ISO 9300 provides the calculation method for the humidity
Publikováno v:
Flow Measurement and Instrumentation. 21:488-496
Many NMIs (national measurement institutes) and calibration laboratories worldwide use a bell prover as a calibration system for gas flow meters. The basic definition and procedure to estimate the bell prover uncertainty have been given in previous s
Publikováno v:
Flow Measurement and Instrumentation. 21:340-346
The purpose of this investigation is to develop flow measuring technology for miniscule flow-rates, in order to calibrate sonic nozzles at lower pressures than the existing pressure ranges, without resorting to an expensive primary standard system. O
Publikováno v:
Flow Measurement and Instrumentation. 20:200-205
Many NMIs (National Measurement Institute) and calibration laboratories are using a mercury sealed piston prover as calibration system for gas flow meter. But it has problems in deciding measuring volume, traveling time of piston and generating low f
Publikováno v:
Journal of Thermal Science. 16:28-32
The mass flow rate measurement using a critical nozzle shows the validity of the inviscid theory, indicating that the discharge coefficient increases and approaches unity as the Reynolds number increases under the ideal gas law. However, when the cri
Publikováno v:
Journal of Thermal Science. 15:355-360
A safety valve functions to control an upper limit of pressure inside the LNG line of transportation. If the pressure inside the safety valve nozzle exceeds a pre-determined value on the valve sheet which plugs the nozzle, an excess of LNG discharges
Publikováno v:
Flow Measurement and Instrumentation. 17:81-86
Recently, critical nozzles have been extensively utilized to measure the mass flow rate in a variety of industrial applications. For the measurement of the mass flow rates in a wide range of operation conditions, the critical nozzle is required to be
Autor:
Se-Jong Jun, Kyung-Am Park
Publikováno v:
Journal of Fluid Machinery. 8:62-66
Autor:
Kyung-Am Park, Hae Man Choi
Publikováno v:
Journal of Fluid Machinery. 7:24-29
The inter-comparison of standards is required to be recognized the calibration results. The sonic nozzle package is developed to do inter-comparison of the bell preyer used to calibrate many gas flow meters for industries. Four national metrology ins