Saudi Pharmaceutical Journal, Vol 28, Iss 1, Pp 74-86 (2020)
The dose emission from DPIs can be affected by the inspiratory parameters achieved by the patient as well as the device in-use. Conventional in-vitro dose emission methodology was used, but instead of using inhalation volume (Vin) of 2 or 4 L and pea