Zobrazeno 1 - 8
of 8
pro vyhledávání: '"J. R. Pekelsky"'
Autor:
J E Decker, J R Pekelsky
Publikováno v:
Metrologia. 34:479-493
Evaluation of the uncertainty for the measurement of a gauge block by optical interferometry is presented. The measurement method which serves as the basis for the uncertainty evaluation is consistent with popular measurement techniques which apply t
Autor:
J R Pekelsky, S. Wingar, Shan Zou, B J Eves, D. Goodchild, Linda J. Johnston, N. Kim, Jennifer E. Decker, David Munoz-Paniagua, Mark T. McDermott, A. Bogdanov
Publikováno v:
Nanotechnology in Construction 3 ISBN: 9783642009792
Nanotechnology emerges out of fundamental science through capability for accurate, repeatable and reproducible measurements on the nanoscale which allows scientists and engineers to accumulate knowledge. Understanding the measurement science is the f
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::c44333b2dceb6204a21cec01d50b7928
https://doi.org/10.1007/978-3-642-00980-8_8
https://doi.org/10.1007/978-3-642-00980-8_8
A laser diffractometer for calibrating grating pitch is being developed at the National Research Council (NRC). Based on the METAS design, it features a precision rotary table, the sample grating mounted in Littrow configuration, a collimated spatial
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::94db41a1c0ce3b76553a338702be93eb
https://nrc-publications.canada.ca/eng/view/object/?id=88041e7b-c9be-4a05-92f5-af5b15f213c2
https://nrc-publications.canada.ca/eng/view/object/?id=88041e7b-c9be-4a05-92f5-af5b15f213c2
Publikováno v:
Journal of Physics: Conference Series. 648:012013
This work describes the development of a grating diffratometer to provide traceable calibration of grating pitch in range 20 μm to 350 nm. The approach is based on the Littrow configuration in which a laser beam is directed onto the grating which is
A long gauge in the form of a thick-walled tube is converted to a Fabry-Perot cavity by attaching partial reflectors to its ends. The length of this gauge is determined from a number of absolute frequency measurements of the cavity resonances. A set
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::b3db405d9cde704a3cf105df0c547bfe
https://nrc-publications.canada.ca/eng/view/object/?id=e47ee6a8-2866-486c-b1c1-cb07da5de434
https://nrc-publications.canada.ca/eng/view/object/?id=e47ee6a8-2866-486c-b1c1-cb07da5de434
Publikováno v:
Image Processing, Analysis, Measurement, and Quality.
For the assessment of human psychophysical and perceptual response to displays, fast and accurate radiometric measurements of display devices are needed. This paper reviews design criteria for spectroradiometers to be used for real-time colour CRT ch
Autor:
R. W. Morrison, J. R. Pekelsky
Publikováno v:
Conference on Electrical Insulation & Dielectric Phenomena - Annual Report 1973.
Induced birefringence, or the electro-optic Kerr effect, has been used by a number of authors to study the steady-state distribution of space charge in a parallel-plate capacitor with nitrobenzene or chlorobenzene.1-9
Publikováno v:
Journal of Physics: Conference Series; Oct2015, Vol. 648 Issue 1, p1-1, 1p