Zobrazeno 1 - 3
of 3
pro vyhledávání: '"G. H. Langner"'
Publikováno v:
Health Physics. 62:87-90
Quality control exposures of commercial alpha-track radon monitors have been conducted approximately weekly at the U.S. Department of Energy (DOE) Grand Junction Projects Office since early 1987 in support of DOE remedial action programs. The results
Publikováno v:
Health physics. 61(4)
New measurements on chemically separated samples of 214Bi have yielded a mean half-life value of 19.71 +/- 0.02 min, where the error quoted is twice the standard deviation of the mean based on 23 decay runs. This result provides strong support for th
Publikováno v:
Health physics. 57(1)
Direct observation of the 218Po alpha-peak decay with a microcomputer-controlled alpha-spectrometer yielded a mean half-life value of 3.040 +/- 0.008 min, where the error quoted represents twice the standard deviation of the means from 38 separate de