Zobrazeno 1 - 8
of 8
pro vyhledávání: '"Felix Wipf"'
Autor:
Aniruddh N. Nayak, Arash Araghi, Beatrice Pavan, Felix Wipf Dipl-Ing, Robert L. Boudreaux, Marc C. Jacofsky, Kellen Worhacz
Publikováno v:
Journal of orthopaedic trauma. 32(12)
OBJECTIVE There are limited biomechanical data supporting the use of anterior or superior-lateral precontoured clavicle plates for the treatment of displaced Neer type II-A clavicle fractures. The objectives of this study were as follows: (a) compare
Autor:
P. B. Robioneck, Peter Augat, Christian Heiss, Felix Wipf, Katrin S. Lips, Ahmed Abdulazim, Reinhard Schnettler, Volker Alt
Publikováno v:
Journal of Biomedical Materials Research Part B: Applied Biomaterials. 104:170-179
Bioresorbable implants may serve as an alternative option for the fixation of bone fractures. Because of their minor inherent mechanical properties and insufficient anchorage within bone bioresorbable implants have so far been limited to mechanically
Autor:
Stefanie Konowalczyk, Felix Wipf, Utku Kandemir, Peter Augat, Geert von Oldenburg, Ulf Schmidt
Publikováno v:
Journal of orthopaedic trauma. 31(8)
Objectives To investigate whether (1) the type of fixation at the shaft (hybrid vs. locking), (2) the position of the plate (offset vs. contact) and (3) the implant material has a significant effect on (a) construct stiffness and (b) fatigue life in
Autor:
Jakob Hallbauer, Sascha Rausch, Thomas Mückley, Andreas Gräfenstein, Claudia Beimel, Felix Wipf, Gunther Olaf Hofmann, Kajetan Klos
Publikováno v:
Foot and Ankle Surgery. 20:180-185
Background Polyaxial locking plates are becoming popular for the fixation of distal fibula fractures. This study establishes how construct stiffness and plate loosening, measured as range of motion, differs between lateral and posterolateral plate lo
Publikováno v:
Journal of Orthopaedic Trauma. 28:130-136
Objective The investigation hypothesized that in current anatomical precontoured plates, angular stability plays only a minor role for the efficacy of the osteosynthesis at the distal humerus. Methods An AO C2.3 fracture model was simulated and osteo
Autor:
Felix Wipf, Mark E. Magill, Thomas J. Moore, Angela S.P. Lin, Chelsea M. Fechter, Robert E. Guldberg
Publikováno v:
Journal of Orthopaedic Surgery and Research
Background Metallic reconstruction plates used for fracture stabilization typically require intraoperative contouring for patient-specific anatomical fit. Despite this, characterization of plate mechanical properties after contouring has previously b
Autor:
Felix Wipf, Nils Reimers, Mauricio Reyes, Serena Bonaretti, Benjamin Kienast, Michel Vallotton, Arndt P. Schulz, Nina Kozic
Publikováno v:
The Open Orthopaedics Journal
Schulz, Arndt P.; Reimers, Nils; Wipf, Felix; Vallotton, Michel; Bonaretti, Serena; Kozic, Nina; Reyes, Mauricio; Kienast, Benjamin J. (2012). Evidence Based Development of a Novel Lateral Fibula Plate (VariAx Fibula) Using a Real CT Bone Data Based Optimization Process During Device Development. Open orthopaedics journal, 6(1), pp. 1-7. Sharjah: Bentham Open 10.2174/1874325001206010001
Schulz, Arndt P.; Reimers, Nils; Wipf, Felix; Vallotton, Michel; Bonaretti, Serena; Kozic, Nina; Reyes, Mauricio; Kienast, Benjamin J. (2012). Evidence Based Development of a Novel Lateral Fibula Plate (VariAx Fibula) Using a Real CT Bone Data Based Optimization Process During Device Development. Open orthopaedics journal, 6(1), pp. 1-7. Sharjah: Bentham Open 10.2174/1874325001206010001
Development of novel implants in orthopaedic trauma surgery is based on limited datasets of cadaver trials or artificial bone models. A method has been developed whereby implants can be constructed in an evidence based method founded on a large anato
Publikováno v:
World journal of orthopedics. 6(9)
AIM: To determine whether use of a precontoured olecranon plate provides adequate fixation to withstand supraphysiologic force in a comminuted olecranon fracture model. METHODS: Five samples of fourth generation composite bones and five samples of fr