Zobrazeno 1 - 5
of 5
pro vyhledávání: '"Ben Ubod"'
Autor:
John A. Ormiston, Ben Ubod, Stephen J. Waite, P Ding, Trine Ørhøj Barkholt, Bruce Webber, Mark Webster
Publikováno v:
Barkholt, T Ø, Ormiston, J A, Ding, P, Webber, B, Ubod, B, Waite, S & Webster, M W 2018, ' Coronary balloon catheter tip damage. A bench study of a clinical problem ', Catheterization and Cardiovascular Interventions, vol. 92, no. 5, pp. 883-889 . https://doi.org/10.1002/ccd.27441
Catheterization and Cardiovascular Interventions
Catheterization and Cardiovascular Interventions
OBJECTIVES: To confirm clinically that coronary balloon catheter tips may be damaged during bifurcation treatment with side-branch access through the side of a stent. On the bench, we aimed to assess the susceptibility of different balloon designs to
Externí odkaz:
https://explore.openaire.eu/search/publication?articleId=doi_dedup___::3d13821e7ad1289384ee110b9c731f4e
https://pure.au.dk/portal/da/publications/coronary-balloon-catheter-tip-damage-a-bench-study-of-a-clinical-problem(8372a818-9457-4252-9bdc-99bfaa44cb74).html
https://pure.au.dk/portal/da/publications/coronary-balloon-catheter-tip-damage-a-bench-study-of-a-clinical-problem(8372a818-9457-4252-9bdc-99bfaa44cb74).html
Publikováno v:
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 10(12)
AIMS Stent fracture is important because it may cause adverse events. The interventional cardiologist needs independent data to aid stent selection. Stent designers need data to improve stent design. We used a repetitive bend test to compare durabili
Publikováno v:
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 10(10)
Aims To provide bench insights which may predict safety and efficacy of side branch dilatation (SB) and kissing balloon post-dilatation (KBPD) in Absorb everolimus-eluting bioresorbable scaffolds deployed in bifurcations. Methods and results Stages o
Publikováno v:
Circulation. Cardiovascular interventions. 7(1)
Background— Stent longitudinal distortion, while infrequent, can lead to adverse clinical events. Our first bench comparison of susceptibility of different stent designs to distortion applied force to the entire circumference of the proximal stent
Publikováno v:
Journal of the American College of Cardiology. 62(18)